Where the liquid actually goes
After a mix, a thaw or a heat cycle, droplets cling to the well walls and the sealing film. The volume in the well is unchanged, but the volume at the bottom is not — and that is the volume the optical system and the pipette see.
Why bubbles matter optically
In qPCR and absorbance reads, a bubble at the meniscus scatters light and shifts the baseline. A short spin collapses the bubble. It is a cheaper fix than rerunning the plate.
Swing-out versus fixed-angle
Swing-out buckets present the plate perpendicular to the centrifugal force, so liquid collects squarely at the bottom of each well. That geometry is why microplate centrifuges are almost always swing-out designs.
Speed is not the goal
Plate spins are about moving a small volume a short distance. A few hundred × g for a few seconds does the job. Running at maximum speed achieves nothing extra and puts more stress on the plate and the buckets.
Balance the load
Two plates opposite each other, or a plate and a balance blank. Modern units detect imbalance and stop, but the stop itself costs you a run. Load symmetrically as a habit rather than relying on the safety cut-out.