Purity is a ratio, not an absolute
HPLC purity is the area of the target peak divided by the total area of all peaks detected at the chosen wavelength. A figure of 99.2% means that 0.8% of the integrated area belongs to something else. It says nothing about what that something is.
The wavelength decides what you see
Most peptide methods run at 214nm or 220nm, where the peptide backbone absorbs. Change to 280nm and you see aromatic residues instead. A lot can look cleaner at 280nm than at 214nm simply because the impurities lack aromatic groups and disappear from view.
Co-elution hides real problems
Two sequences with similar hydrophobicity can elute together and integrate as one clean peak. This is why mass spectrometry confirmation matters alongside the chromatogram — the mass tells you what the peak is, the chromatogram tells you how much of it there is.
Compare figures from the same method
A 98% result from a method with a shallow gradient and a long run time is not the same as 98% from a fast 10-minute method. When comparing suppliers, compare the method as well as the number.
Ask for the chromatogram
A single figure is a claim; the chromatogram is the evidence. Suppliers who can produce the trace for the specific lot are showing you the working, not just the answer.