Readability is not accuracy
A balance that displays 0.1mg does not necessarily measure to 0.1mg across its range. Readability is the smallest increment the display shows; accuracy is how close that display is to the true mass. The specification that matters for small samples is repeatability at the low end of the range.
The 220g / 0.1mg combination
A semi-micro balance with a 220g capacity covers both a small powder aliquot and a heavier container without switching instruments. That combination is why semi-micro designs have become the default in laboratories that weigh a bit of everything.
Static is the dominant error source
Below about 20mg, electrostatic charge can move a reading by tenths of a milligram. Dry powders and plastic weigh boats are the usual culprits. A built-in ioniser neutralises the sample and container before the reading settles, which is the single most effective fix.
Draft and vibration matter more at five places
A large glass draft shield is not cosmetic. Air currents from a doorway or a nearby fume hood will show up as a wandering last digit. Put the balance on a solid bench away from traffic before blaming the instrument.
When a four-place balance is enough
If your smallest sample is above 50mg and your tolerance is a few tenths of a milligram, an analytical balance will serve you well and cost less. The five-place instrument earns its price when the sample itself is small.