What static does to a reading
A charged powder repels or attracts the pan, and the resulting force is read as mass. The error can be positive or negative, and it changes as the charge dissipates — which is why the display keeps moving after you think the reading should be stable.
Why lyophilised powders are worst
Freeze-dried material has an enormous surface area and very low moisture, both of which favour charge accumulation. Transferring it from a plastic container or across a dry room makes it worse.
Ionisers neutralise, they do not blow
A fanless ioniser produces positive and negative ions that neutralise the charge on the sample and container. Older units used a fan to distribute ions, which introduced its own draft problem. Fanless designs avoid that trade-off.
Practical habits that help
Use metal or anti-static weigh boats rather than plastic where possible. Let the sample sit in the ioniser's field for a few seconds before recording the reading. Keep the draft shield closed. Avoid weighing immediately after moving the container across a dry bench.
When to suspect something else
If an ioniser and good technique still leave you with unstable readings, check for vibration and air currents before assuming the sensor is failing. A drift that appears only when the fume hood is running is a draft problem.